DataMaster Error Codes
Please Note: These definitions were placed on this site for informational purposes only. Seeing one of these errors does not necessarily mean a test is invalid, or that the instrument required a repair. Please contact your local breath test technician to clarify any questions you may have.
|
Database |
Display |
Definition |
|
1 |
System Won’t Zero |
Unable to zero detector voltage. |
|
2 |
Temperature Low |
Sample chamber temperature falls to 45 degrees C or below. |
|
3 |
Temperature High |
Sample chamber temperature rises to 55 degrees or above. |
|
5 |
Radio Interference |
Radio frequencies detected. |
| 6 | Fatal System Error | Instruments internal memory is not responding properly. |
|
7 |
Calibration Error |
Internal standard does not read within 10% of the value determined at the time of calibration. |
|
8 |
Printer Error |
Printer not responding properly. |
|
9 |
Ram Error (Addresses) |
RAM checksum does not match that calculated following the last write. |
|
10 |
Pump Error |
Flow detector does not detect pump operation. (pump not drawing enough flow) |
|
11 |
Blank Error |
Instrument detects a reading greater than .001 during the blank test. |
|
12 |
Detector Overflow |
Detector output exceeds the 1.999 volts that is readable by the instruments analog/digital converter. (Too much of the infrared light is being absorbed) |
|
13 |
Filter Error |
Filter solenoid not activating properly. (filter not going into place as designed) |
|
15 |
Simulator Out of Range |
Simulator reading outside acceptable limits of .072-.088. |
|
17 |
Data Memory Battery Low |
RAM Battery backup failing. |
|
19 |
Ambient Fail |
Ethanol or other substance detected in sample chamber after purge. (Instrument unable to completely purge sample chamber to .003 or less.) |
|
20 |
Samples outside 10% |
Two subject samples do not fall within +/-10% of their mean. |
|
V |
Invalid Sample |
Instrument unable to determine if mouth alcohol or subject blowing improperly. |
|
X |
Interferant |
Acetone or similar molecular structure to acetone detected in sample. |
|
R |
Refusal |
Subject refused to provide breath sample, refusal entered by operator. |
|
I |
Incomplete Test |
Subject did not provide breath sample within 2 minute frame allotted by the instrument |
|